Oil in Water Transmitter
SS311-01-LORA
Oil in Water Transmitter
Product Highlights
● Fluorescence sensing with light-filtering algorithm.
● Automatic temperature compensation.
● LoRa spread-spectrum wireless transmission.
● IP68 probe and IP65 transmitter housing.
● Gateway-based networking and Bluetooth configuration.
● Automatic temperature compensation.
● LoRa spread-spectrum wireless transmission.
● IP68 probe and IP65 transmitter housing.
● Gateway-based networking and Bluetooth configuration.
Product Overview
Overview
• Measures oil-related fluorescence in water for pollution monitoring and early warning.
• Relative fluorescence units support analysis of the relationship between signal intensity and oil concentration.
• Sends readings by LoRa to a compatible gateway for centralized platform monitoring.
| Item | Specification |
| Model | SS311-01-LORA |
| Measurement principle | Fluorescence |
| Measurement range | 0–50 ppm; 0–100 RFU |
| Measurement error | ±5% FS at 25 °C; temperature: ±0.5 °C |
| Resolution | 0.01 ppm; 0.1 RFU; temperature: 0.1 °C |
| Response time | ≤30 s, T90 |
| Probe operating temperature | 0–40 °C |
| Housing environment | −40 to +60 °C; 0–95% RH, non-condensing |
| Power supply / average power | DC 10–30 V / ≤0.7 W |
| Data transmission | LoRa proprietary wireless protocol; RS485 Modbus-RTU interface |
| RS485 baud rate | 4800 bit/s default; 1200, 2400, 4800, 9600, 19200, 38400, 57600 or 115200 bit/s |
| Data upload interval | 300 s default. Set multiples of 60 s; other values are delayed to a 60 s multiple. |
| Ingress protection | Probe: IP68; transmitter housing: IP65 |
| Probe cable length | 5 m standard |
| Housing material | Corrosion-resistant material |
| Housing dimensions | 117 × 87 × 43 mm maximum |
| Probe dimensions / thread | 147 mm dimension shown; Ø30 mm; NPT 3/4 thread |
| LoRa channels | Transmit and receive channels: 1–64 |
| Spreading factor | 7–9; default 9; must match the paired host |
| Time-slot interval | 3000 ms default, read-only |
| Collector Modbus master baud rate | 4800 bit/s default; 2400, 4800 or 9600 bit/s; match the connected RS485 device |